Thank you for taking an interest in Mallinckrodt Sensor Systems Inc part number N0-059, which is currently available for purchase today here on ASAP Aerospace. You have reached our Request for Quote (RFQ) page for this part number N0-059, where we make it easy to secure competitive pricing and procurement options from our staff.
Part Number N0-059 Details:
Part number N0-059 traces back to Mallinckrodt Sensor Systems Inc (CAGE Code 0CNN1), having been classified under NSN 6630013648555 and item name Analyzer Blood Gas. With its NSN, this part number belongs to FSG 66 Instruments and Laboratory Equipment and FSC 6630 Chemical Analysis Instruments. Additional listing data derived from the NSN includes the NIIN 013648555, the INC 38049, and NCB code 01 (USA).
Note : We Will Not Share Your Information To Any Third Parties.
MRC | Criteria | Characteristic |
---|---|---|
DPTH | Depth | 18.300 INCHES |
ELEC | Voltage In Volts | 120.0 AND 230.0 |
MRC | Decoded Requirement | Clear Text Reply |
CBBL | Features Provided | DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION |
CXCY | Part Name Assigned By Controlling Agency | BLOOD GAS ANALYZER 120/230V 50/60 HZ AC: |
ACDC | Current Type | AC |
CYFV | Method Of Analysis | POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
ACZB | Frequency Rating | 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
CYFY | Analysis Test/Measurement Range | PO2 0 TO 600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MMOL/L; NA 100 TO 200 MMOL/L; CA 0.25 TO 4.25 MMOL/L; HCT 10 TO 75 PCT; HC03 3.0 60 MMOL/L TCO2 3.0 TO 60 MMOL/L; BASE EXCESS -30 TO +30 MMOL/L; O2 SAT 30 TO 90 PCT. |
CYHM | Test Sample/Specimen Type | WHOLE BLOOD |
ABRN | Height | 8.400 INCHES |
WDTH | Width | 9.100 INCHES |
AQXY | Test Type For Which Designed | BLOOD GASES AND ELECTROLYTES,HC03,HCT,BASE EXCESS AND 02 SAT |
CYFZ | Test Sample/Specimen Size | 0.5 ML |
FEAT | Special Features | UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD, CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT. |